Biosafety Cabinet Class III – 4 FT
The Biosafety Cabinet Class III – 4 FT is a fully enclosed, gas-tight biological safety cabinet designed for maximum personnel, product, and environmental protection while handling highly infectious and hazardous biological agents. Featuring a negative pressure operating system, double H14 HEPA filtration, integrated glove ports, side pass box, and 100% exhausted airflow, this cabinet provides the highest level of biological containment for BSL-3 and BSL-4 laboratory applications. Built with SS304 stainless steel construction and an LCD controller, it delivers exceptional safety, durability, and reliable performance.
- Estimated Delivery : Up to 4 business days
- Free Shipping & Returns : On all orders over $200
Applications
- BSL-3 & BSL-4 Laboratories
- High-Risk Microbiology Laboratories
- Virology Laboratories
- Pharmaceutical & Biotechnology Research
- Oncology & Cytotoxic Drug Preparation
- Genetic Research & Recombinant DNA Work
- Animal Research Facilities
- Advanced Biological Containment Laboratories
Key Features
- Class III maximum containment biosafety cabinet
- Fully enclosed gas-tight construction
- Negative pressure operating system
- 100% exhausted airflow with no air recirculation
- Dual H14 HEPA filtration system (Supply & Exhaust)
- 99.997% HEPA filter efficiency at 0.3 µm
- Integrated ergonomic glove ports
- Side pass box with interlock system
- SS304 stainless steel work chamber
- UV sterilization lamp
- LED work area illumination
- Airtight internal electrical sockets
- LCD digital controller
- Easy-to-clean stainless steel interior
- Designed for maximum operator and environmental safety
Technical Specifications
- Model: BSC-SS-III-4F
- Cabinet Type: Class III Biosafety Cabinet
- Size: 4 FT
- External Dimensions (W × D × H): 1372 × 838 × 2032 mm
- Internal Dimensions (W × D × H): 1200 × 600 × 700 mm
- Work Surface Height: 750 mm
- Side Pass Box Size (W × D × H): 457 × 457 × 457 mm
- Air Inflow Velocity: 0.40 m/s (80 fpm)
- Downflow Velocity: 0.30 m/s (60 fpm)
- Negative Pressure: -120 Pa (or as specified)
- Supply HEPA Filter: H14, 99.997% @ 0.3 µm
- Exhaust HEPA Filter: Dual H14, 99.997% @ 0.3 µm
- Exhaust Air: 100% Exhausted
- Noise Level: ≤65 dB(A)
- Illumination: ≥1000 Lux
- UV Lamp: 30W
- LED Lamp: 24W
- Work Area Material: SS304 Stainless Steel
- Front Window: Laminated Safety Glass
- Power Supply: 220V, 50Hz
- Power Consumption: 850W
- Controller: LCD Controller
Suggested Price